Canivet Parvovirus-Coronavirus-Giardia Lamblia Ag Tri-Test

Canivet Parvovirus-Coronavirus-Giardia Lamblia Ag Tri-Test

Model No:VD184Q
Abbr.:CaPCG
Assay Time: 5-10 minutes
Specimen: canine feces

Description

 

The Canivet Parvovirus-Coronavirus-Giardia lamblia Ag Tri-Test enables the rapid differential diagnosis of these pathogens by detecting antigens of canine parvovirus(CPV), canine coronavirus(CCV), Giardia lamblia(GIA) in canine feces. It offers two options for result reading: visual judgment or semi-quantitative analysis with the FIC-Q100N Dry Immunoassay Analyzer.

Specification

Intended Use: rapid differential diagnosis of these pathogens by detecting antigens of canine parvovirus(CPV), canine coronavirus(CCV), Giardia lamblia(GIA).

Principle: lateral flow immunochromatographic assay.

Specimen: canine feces.

Materials provided:

10 CaPCG Ag test cards
10 Vials of assay buffer
10 Disposable droppers
10 Specimen collection swabs
1 Instruction for use

Special Features:

High sensitivity and Accuracy.

VD184Q CaPCG

 

Test Procedure

 

Test procedures of visual interpretation

CaPCG

1. Remove the test card from the foil pouch and place it flat on a work surface.
2. Transfer 4 drops of the treated sample from the assay buffer tube into the sample well.
3. Start the timer. Wait for 10 minutes to read the result.

 

Interpretation of visual interpretation results

CaPCG

-Positive (+):

Canine parvovirus virus positive, C line and P line appear.

Canine coronavirus positive, C line and C line appear.

Giardia lamblia positive, C line and G line appear.

[NOTE] Mixed infection: If two or three Test lines (P, C, and G) appear together with the C line, it indicates co-infection.
Example: C line + P line + C line positive indicates co-infection with canine parvovirus and canine coronavirus

-Negative(-): The C line is visible, but the P, C and G line is not visible.
-Invalid: No line appears in the C zone.

 

Test procedures of instrumental interpretation (Standard Test)

CaPCG

1. Turn on the FIC-Q100N Dry Immunoassay Analyzer.
2. Use the analyzer's scanner to scan the QR code on the inner label of the box or the label of bag to import the calibration curve into the analyzer.
3. Remove the test card from the foil pouch.
4. Position the front of the card facing up, ensuring that the sample well is oriented towards the outside of the analyzer's test card slot.
5. Insert the card into the test card slot of the analyzer.
6. Transfer 4 drops of the treated sample from the assay buffer tube into the sample well of the test card.
7. Push the test card to the farthest end of the slot.
8. Tap "Standard Test" on the screen to start the testing. Please refer to the analyzer's operation manual for detailed instructions

 

Test procedures of instrumental interpretation (Quick Test)

CaPCG

1. Turn on the FIC-Q100N Dry Immunoassay Analyzer .
2. Use the analyzer's scanner to scan the QR code on the inner label of the box or the label of bag to import the calibration curve into the analyzer.
3. Remove the test card from the foil pouch.
4. Transfer 4 drops (using a disposable dropper) of the treated sample into the sample well.
5. Wait for 10 minutes, then push the test card to the farthest end of the test card slot.
6. Tap "Quick Test" on the screen to start the testing. Please refer to the analyzer's operation manual for detailed instructions

 

 

 

Order Information

Product No. Product Name Storage Temperature Package Unit
VD184Q Canivet Parvovirus-Coronavirus-Giardia lamblia Ag Tri-Test 2-30°C 10 Tests/Box

 

 

 

 