Feline Calicivirus-Herpesvirus Typel-Panleukopenia Virus Antigen Test

Feline Calicivirus-Herpesvirus Typel-Panleukopenia Virus Antigen Test

Model No:VD079
Abbr.:FCV-FHV-FPV Ag
Assay Time:5-10 minutes
Specimen:FCV-FHV Ag--secretions from cat's eyes,nsal cavities,and anus FPV Ag---feces

Description

Specification

 

Feline Calicivirus-Herpesvirus Typel-Panleukopenia Virus Antigen Test

Intended Use: for the differential diagnosis of Feline Calicivirus virus antigen (FCV Ag), Feline Herpesvirus Type-1 antigen (FHV Ag) and Feline Panleukopenia virus antigen (FPV Ag).

Principle: lateral flow immunochromatographic assay.

Specimen: FCV-FHV Ag--- secretions from cat's eyes, nasal cavities, and anus.

                  FPV Ag--- feces or vomit.

Materials provided:

10 test pouches with cards

10 vials of assay buffer FPV Ag

10 vials of assay buffer FHV-FCV Ag

20 disposable droppers

20 bags of swab sticks

1 package insert

Special Features:

Simultaneous detection of FCV, FHV and FPV.

One-step testing procedure : Quick & Accurate.

High sensitivity and specificity.

 

product-602-659

product-589-176

product-584-69

FCV-FHV-FPV Ag

TEST PROCEDURE

Allow all materials, including specimen and test device, recover to 15-25℃ before running the assay.

FCV-FHV Ag test procedure

Collect cat's ocular, nasal or anus secretions with the swab stick and make the swab wet sufficiently.

Insert the swab into the provided assay buffer tube. Agitates it to get efficient sample extraction.

Take out the test device from the foil pouch and place it horizontally.

Suck the treated sample extraction from the assay buffer tube and place 3 drops into the sample hole "S" of the test device.

Interpret the result in 5-10 minutes. Result after 10 minutes is considered as invalid.

product-907-364

FPV Ag test procedure

Collect cat's fresh feces or vomit with the swab from cat's anus or from the ground.

Insert the swab into the provided assay buffer tube. Agitates it to get efficient sample extraction.

Take out the test device from the foil pouch and place it horizontally.

Suck the treated sample extraction from the assay buffer tube and place 4 drops into the sample hole "S" of the test device.

Note: If the liquid will not flow through the surface of the test strip in 30 seconds, please add another drop of the treated sample extraction.

Interpret the result in 5-10 minutes. Result after 10 minutes is considered as invalid.

product-907-364

INTERPRETATION OF RESULTS

Positive (+): The presence of both "C" line and zone "T" line, no matter T line is clear or vague.

Negative (-): Only clear C line appear. No T line.

Invalid: No colored line appears in C zone. No matter if T line appears.

 

Order Information

 

Product No. Product Name Storage Temperature Package Unit
VD079

Feline Calicivirus-Herpesvirus Typel-Panleukopenia Virus Antigen Test

2-30°C 10 Tests/Kit

Hot Tags: feline calicivirus-herpesvirus typel-panleukopenia virus antigen test, China feline calicivirus-herpesvirus typel-panleukopenia virus antigen test suppliers, manufacturers, factory, Feline Leukemia Virus Antigen FELV Ag Test, Fpv Ag Test Kit, Fiv Ab Felv Ag Test, Feline Giardia Ag Test Kit, Feline Calicivirus Antigen FCV Ag Test, Feline Toxo Ab Test

(0/10)

clearall